Free Thyroxine ( T4 ) RIA by equilibrium dialysis (FT4ED)

Free Thyroxine Hormone Ria kit by Equilibrium Dialysis using QD-CELL™ Disposable Quantitative Dialysis Cell.

The Free thyroxine (FT4) by Equilibrium Dialysis (FT4ED) RIA Kit is a radioimmunoassay for measurement of free Thyroxine (T4) after dialysis. The dialysis is performed using a QD-CELL™ Disposable Quantitative Dialysis Cell.
Sample Preparation:
200 µL of a sample and 1350 µL of dialysis buffer are added to a sample and buffer compartment. The serum sample is allowed to dialyze for 17 hours at 37°C. 700µL of dialysate is then used in an RIA procedure for determining the concentration of free T4.

Kit components:
FT4ED kit DFT41000
First Antibody (Anti T4) DFT41001
RIA buffer DFT41002
Dialysis buffer DFT41003
Second antibody DFT41004
T4 Standard 12.8ng/dL DFT41005
Tracer DFT41006






Sensitivity:

Limit of detection <0.1 ng/dL
Limit of quantification 0.2 ng/dL
Analytical measurement range

0.2 – 12.8 ng/dL
Procedure overview:

  1. Overnight dialysis of samples
  2. One hour incubation of dialysate, standards, and controls with anti-thyroxine and I125-labelled thyroxin
  3. Thirty minute incubation on ice
  4. One hour incubation with second antibody
  5. Centrifuge, decant then count